Purpose: To align the mask to the layers of processing under the photoresist or to simply expose resist through a mask
Location: Pettit and Marcus Cleanrooms
Tools: Karl Suss MA6 Mask Aligner - Pettit, Karl Suss TSA MA-6 Mask Aligner - Marcus
Supplies Needed: mask, mask holder, correct size wafer chuck, tweezers, blue tape (if necessary)
Setup Procedures:
- Choose the proper exposure wavelength on the mask aligner tool and the intensity measurement tool.
- Place the sensor of the intensity measurement tool in the center of the mask aligner chuck
- If possible, measure the intensity through the mask.
- Place the mask on top of the sensor so that dark space does not cover the sensor
- If not enough clearfield space, measure with the sensor alone
- Start a lamp test and not the intensity reading
- Calculate the exposure time by taking the dosage found on the photoresist data sheet and dividing by the measured intensity.
- Be sure to subtract 1.0 mW/cm2 from the measured intensity to account for loss from the mask if not able to measure through the mask
- Program the calculated time into the recipe on the mask aligner
Number of Substrates: 1
Process Procedures:
- Load your mask onto the properly sized mask holder with the chrome side up, and then turn on the mask vacuum
- Load the mask holder into the mask aligner. The chrome side should now be facing down.
- Choose the mask aligner chuck so that you sample will not hang over the edges
- Load the mask aligner chuck
- Load your sample onto the chuck
- If your sample does not cover the entire chuck, use blue tape to cover the rest
- Cut a hole in the blue tape slightly smaller than your sample size
- Cut the blue tape on a counter or table away from the mask aligner
- Do not cut on the chuck
- Do not cut the vacuum bladder for the vacuum chucks
- Load your sample into the mask aligner
- Align your sample as needed and expose
- Unload your sample
- Transfer you sample to the hotplate or oven for a post-exposure back, if needed.
- Unload your mask